Item 70-S12Z00 or 70-S12Z10
Steel & Concrete Test Frame
- Configuration
- 500/1000 kN

Comprehends high accuracy load cell, tensile jaws with wedges grips and 150 mm travel displacement transducer allowing test execution under crosshead separation control to EN EN ISO 6892-1 (method B) and EN 15630-1, when connected to our…
MeetsEN 15630-1 · ISO 6892-1
Technical specifications
| Load capacity in tension | 500 kN |
|---|---|
| Load capacity in compression * | 1000 kN |
| tensile jaws including | 4 wedge grips for flats up to 13 mm thickness, 4 wedge grips for rounds up to 26 mm dia. and 2 sets of grips’ liners 4 and 8 mm thick. |
| Maximum distance between grips (tensile mode) | approx. 300 mm |
| Specimen length (tensile mode) | approx. 500 mm |
| Max vertical clearance for compression (with 70-S0012/1)* | 695mm |
| Max. ram travel | 150 mm |
| Distance between columns | 310 mm |
| Overall dimensions approx (w x d x h). | 605 x 430x 1750 mm |
| Weight approx. | 550 kg |
| For model | 70-S12Z00 only |
Technical notes
- Frame
Features
- For tensile tests up to 500 kN for testing steel rebars up to 26 mm dia.
- Connectable as additional frame to AUTOMAX Multitest control console
- Vertical daylight 450 mm
- Piston travel 150 mm
- Compatible specimens: max. rebar dia. 26 mm – max. flat specimen thickness 13 mm
- Heavy duty high-functionality jaws rated for severe prolonged
- For tensile tests only it’s also available a dedicated version fitted with high accuracy load cell
Included
| a set of tensile holders, 4 wedge grips for flats up to 13 mm thickness, 4 wedge grips for rounds up to 26 mm dia |
Required equipment
| Spherical seat and compression platens for concrete specimens |
Accessories
| — | It is also available a version for tensile tests only fitted with high accuracy load cell. |
|---|---|
| 70-S0012/1 | Set of lower platen and upper platen with spherical seat, 216 mm dia., for testing cylinders up to dia. 160×320 mm and cubes from 100 to 150 mm. To be completed with the suitable distance pieces depending on the specimen height. Weight approx.: 45 kg |
| — | Distance pieces Made of steel, these pieces are used to reduce the vertical clearance of the compression machine to a height that is appropriate for the size of the specimen and considering that, in general, the maximum piston travel is 150mm. |
| Code | Dimensions (diameter xheight, mm) |
| 50-C9080 | 200 x 30 |
| 50-C9082 | 200 x 50 |
| 50-C9083 | 200 x 68 |
| 50-C9084 | 96 x 158 |
| 50-C9086 | 200 x 100 |
| 50-C9087 | 96 x 130 |
Show 2 more items
| 70-S0012/2 | Front and rear fragment guard |
|---|---|
| 70-C0961/H | Coaxial electronic extensometer for round specimens from 6 to 26 mm dia. Gauge length: 200 mm. Travel: 50 mm. |