TerraField logo
Item TF-1PFLTAI

Dynamic Triaxial System Dynatriax EmS

Dynamic Triaxial System Dynatriax EmS

DYNATRIAX EmS benefits from CONTROLS Group new EmS “Electromechanical Servoactuation” technology.

MeetsAASHTO T307
In stock
Price: request quote
Add to quote

Technical specifications

Technical specifications for Dynamic Triaxial System Dynatriax EmS
Maximum dynamic load15 kN
Maximum static load10 kN
Maximum vertical travel50 mm (different travels available on request)
Maximum testing frequencymore than 10 Hz (depending on testing conditions)
Volume change measurement100 cc volume change device with automatic flow inversion
Max sample diameter100 mm
Maximum confining pressure1000 kPa
Maximum back pressure1000 kPa
Close loop control frequency10 kHz
Input channels16 – Real resolution 16 bits
Voltage208-230 V, 50-60 Hz, 1 ph or 110 V, 60 Hz, 1 ph

Features

  • Electromechanical Servoactuation, no need of compressed air or hydraulic power supply for the vertical force
  • Capability to perform Static (effective stress and stress path), Dynamic soil triaxial tests
  • Maximum dynamic load: 15 kN
  • Maximum static load: 10 kN
  • Triple-axes closed loop control of axial load/displacement, cell and back pressure
  • Automatic compensation of cell/back pressure during dynamic stage
  • Operating frequency more than 10 Hz (depending on test conditions)
  • Complete automation of all test stages using an high sensitivity closed loop PID feedback
Show 9 more features
  • Options available for bender elements testing and for local strain transducers
  • Standard and user defuned wave shapes which can mimic the real in situ measurements (earthquakes)
  • Test set-up by unique programmable multi-stage procedure. When a test is running, it is possible to access all parameters to expand and modify the stages as per the response of the specimen
  • Manual and automatic emergency shut off function
  • Multitasking, user-friendly Windows-based software pre-installed on the PC
  • Extra-accurate transducers calibration adopting linear or polynomial regression or multi-coefficient linearization
  • Axial load/displacement
  • Cell pressure up to 1000 kPa
  • Back pressure up to 1000 kPa

Included

Included for Dynamic Triaxial System Dynatriax EmS
Reaction frame
Robust 2-columns construction, stiff and compact
Electromechanical vertical load application
Data acquisition, process and control
high performance motorized actuator, 15 kN capacity, backlash free and noiseless
sophisticated PID closed-loop control, ensuring load is reached fast, smoothly and accurately and then maintained with high level of accuracy. The submersible load cell delivers high accuracy from the lowest values
The CDC manages up to 3 closed loop axes (axial load/displacement, cell and back pressure) with an effective loop rate of 10 kHz and performs the test completely automatically including the on/off valves for the drainage line and for the air supply to the triaxial cell. It is linked to the PC via LAN/Ethernet link (1 Gbit/s) and comes with 16 transducer ports (real resolution 16 bit).
PID control
An extremely efficient algorithm with larger gain ranges gives enhanced sensitivity, making it easier to tune the system and achieving a more accurate wave shapes. Recently added features include manual and automatic amplitude control which compensate for small changes that may occupr in the system during cycling, ensuring that the required peaks are consistently reached. The improved tuning panel, with its more user-friendly interface, provdes all the tools necessary to optimize the system control during static and cyclic test stages

Ordering information

Ordering information for Dynamic Triaxial System Dynatriax EmS
31-WF70E06Dynatriax EmS – Automatically PC controlled dynamic triaxial system with 15kN electromechanical servoactuation. Supplied complete with PC, CDCtriax control and 16channels data acquisition unit and software. 208-230V,50-60Hz,1ph.

Accessories

Accessories for Dynamic Triaxial System Dynatriax EmS
31-WF711925kN capacity submersible load cell to be used with Dynatriax (dynamic triaxial system)
31-WF711810kN capacity submersible load cell to be used with Dynatriax (dynamic triaxial system)
31-WF71175kN capacity submersible load cell to be used withDynatriax (dynamic triaxial system)
31-WF7121Axial displacement transducer +/-25mm to be used with Dynatriax (Dynamic Triaxial system)
29-WF4412Automatic volume change apparatus allowing remote flow control. Compatible with WF automatic Static/Cyclic triaxial systems AUTOTRIAX/DYNATRIAX
28-WF6300/APressure transducer 0-1000 kPa, not compatible with DYNATRIAX ( see model 28-WF6300)
28-WF6310De airing block suitable for banded triaxial cells
31-WF7000/RESUpgrading package for Dynatriax EmS for resilient modulus determination, including software to AASHTO T307, two displacement transducers and mounting braket to fit the transducers to the triaxial cell.